Tools and scripts for evaluating image quality in neutron imaging
The contrast sample from the previous iteration of the project (Kaestner, et al., 2013) has been updated by replacing the PE inset with an inset of Ti. This was done to reduce the effect of starvation streaks that appears in particular for combinations of the high attenuation insets. The contrast sample is intended verify the ability to quantitatively determine the attenuation coefficients for five materials using computed tomography.
The experiment requires the acquisition of a CT projection data set and the following reconstruction to provide the data required for the analysis.
This is CT experiment that requires many projections and the exposure time per projection depends on the pixel size. Some examples based on a 2k x 2k detector would be 625 projections with between 1s and 60s exposure time each. This would result in an experiment time of between 1.5h and 12h including setup. Most of this time doesn’t require staff presence.
The concept of determining the resolution from an edge is in general described in standard documents like ISO 12233 and textbooks (Burke, 1996). For neutron imaging the highest contrast is obtained using gadolinium. The edge device shown in figure 1 has two edges to measure the resolution horizontally and vertically. The frame makes it easy to place the test object on the sample table.
Edge device to determine the edge unsharpness of the imaging system.
This experiment requires three radiographs as a minimum. This would take about 15 minutes to obtain. It is however recommended to make multiple radiographs for each type of image. The total time would be about 1h, staff is required during this time.
The impact of geometric unsharpness is explored by moving the edge device away from the detector. There are two main components in the unsharpness; 1) detector unsharpness and 2) geometric unsharpness. The detector unsharpness is dominant for images acquired with the sample close to the detector, while the geometric unsharpness dominates for samples remotely from the detector. It is possible to measure the impact of geometric unsharpness by moving the edge device away from the detector and making radiographs on each position.
This experiment requires as a minimum one radiograph per distance and reference images for flat field and dark current. The total time would be about 2h depending on how fast the sample positioning can be done, staff is required throughout this experiment.
The resolution of a CT image is more complicated to determine than for the radiographs. First, a complete CT scan must be performed. Second, the material of the test object must allow some neutrons to penetrate the object, otherwise it will not be possible to reconstruct the projections into a valid volume. Also, CT provides volume information, therefore we also need to measure the resolution outside the plane of the central slice. The best shape for this is the sphere. We chose to use a 20 mm copper ball as test object as the curvature is sufficiently low to locally approximate the surface as flat for the edge extraction.
This is CT experiment that requires many projections and the exposure time per projection depends on the pixel size. Some examples based on a 2k x 2k detector would be 625 projections with between 1s and 60s exposure time each. This would result in an experiment time of between 1.5h and 12h including setup. Most of this time doesn’t require staff presence.
A three-dimensional analogue to line pair test devices for radiography is to use a set of vials filled with small balls. Each vial is filled with single diameter balls. The resulting packing contains voids and highly attenuating material (we have chosen to use copper). The four vials in the sample box are filled with ball diameters 0.5, 0.8, 1.0, and 2.0 mm. By dividing the ball diameter by the pixel size, you will be able to tell how well the spheres will be represented at the current resolution. In general, you should aim for spheres with diameters greater than ten pixels, figure 3. With smaller spheres the system blurring will smooth the pore space making it hard to uniquely identify the edges.
The sphere material copper has an attenuation coefficient of 1.07 1/cm for thermal neutrons. This allows to arrange the samples as in figure 4, or piled on top of each other. This resolution sample is relatively small. Therefore, you can even include it in your experiments to provide a resolution reference in your experiment data. The arrangement of the vials has an impact on the number of projections as a single vial is less wide than all four at once.
This is CT experiment that requires many projections and the exposure time per projection depends on the pixel size. Some examples based on a 2k x 2k detector would be 625 projections with between 1s and 60s exposure time each. This would result in an experiment time of between 1.5h and 12h including setup. Most of this time doesn’t require staff presence.
All experiments should be possible to complete within two days of beam. For weaker neutron sources this may be longer.